Building of an experimental set-up of laser induced plasma spectrometer

Xu Guowei,Lu Xianyang,Fei Teng,Du Xuewei,Wang Shengbo,Wang Qiuping
DOI: https://doi.org/10.3969/j.issn.1001-232X.2010.04.002
2010-01-01
Abstract:Laser induced plasma spectrometer(LIPS)is an instrument for element analysis with the advantages of real-time,on-line,non-contacting and multi-element detection.It doesn't need sample pretreatment,and can do rapid on-line analysis of elements in iron and steel samples.We have built an experimental set-up of laser induced plasma spectrometer,working in the range of 235~260nm with the spectral resolution up to 0.1nm,by putting a linear array CCD on a Rowland circle.The instrument is mainly used for the detection of carbon in iron and steel samples.Experimental results show that the building of this experimental set-up is successful.