Determination of Micro Tin in Industrial Wastewater by Flow Injection Analysis Integrated with Dual-beam Spectrophotometry
JIN Xu-dong,XU Xiao-ming,PANG Yu-ting,YANG Zhao-yi,ZHAO Zhen-yi
DOI: https://doi.org/10.3969/j.issn.1000-5846.2013.03.009
2013-01-01
Abstract:A novel faster and higher sensitive flow injection derivative spectrophotometry detection technology is proposed,that is,by double beam spectrophotometer in the sample cuvettes and the reference,scan time delay( Δt) produces the first derivative spectra( ΔA / Δt≈dA / dt) for quantitative analysis. While cetyltrimethylammonium bromide ( CTAB ) co-existing,tin and 9- phenyl-2,3,7-trihydroxy-6-fluorine generates colored coordination compound at 510 nm,w hich has the characteristics of absorption under acidic condition. The sample w hich w as injected into a set of self-developed FIA system produced a first derivative spectrophotometry w ith tw o positive peaks,a negative peak and four inflexion points( ΔA1 ,ΔA2 ,ΔA3 ,ΔA4 ) at the same time,w hich w ere all proportional to the content of tin. Experimental results show ed that coexist ion did not affect the color reaction of tin in the presence of tartaric acid,the sensitivity is 1. 8 times higher than normal method and a good selectivity w as maintained in this method. Additionally,the sensitivity approached 4. 8 times by using the plus technology( ΔA = ΔA1 + ΔA2 + ΔA3 + ΔA4 ) .