Characteristics of current of injury at the time of fixation of active fixation pacing leads

XU Xun-long,SU Yan-gang,WANG Shuai,GE Jun-bo
DOI: https://doi.org/10.7695/zgxzqb201306009
2013-01-01
Abstract:Objective To discuss the characteristics and length of time of current of injury( COI) after adequate placement of active-fixation pacing leads( AFPL) for better clinical use. Methods Fifty-nine patients undergoing active-fixation lead implantation were enrolled. R wave amplitude,COI( maximum and at 80 ms) measured by a Medtronic pacing system analyzer( model 2290) from baseline to 10 min after AFPL implantation were recorded. Results Right ventricular lead positioning was attempted 66 times; 59 positions demonstrated good fixation and adequate threshold compared to 7 positions demonstrated high threshold or acute lead dislodgement. 55 sites of good fixation( 93. 2%) showed adequate COI at 1 minute after fixation of AFPL and lasted for 5 minutes in 54 sites( 91. 5%) and for 10 minutes in 50 sites( 84. 7%). while site of high threshold showed no COI and ST 80( the amplitude 80 ms after onset of the R wave deflection) ≤0 mV at 1 minute after fixation of AFPL( P < 0. 001). Conclusion The development of adequate COI within 1 minute after fixation of AFPL can last 10 minutes when the AFPL demonstrated good fixation and had satisfactory threshold. Conversely,no COI or ST 80 ≤0 mV appeared 1 minute after fixation of AFPL as the AFPL showed high threshold or acute lead dislodgement.