Performance Analysis of Bio-Chemical Sensor System Based on Micro-Cantilever

黄渊,陈雁云,刘红,李凯,张青川,伍小平
2008-01-01
Abstract:In a micro-cantilever bio-chemical sensor system based on optical lever,the change of liquid refractive index is able to cause the displacement of spot moving on PSD,which will affect the system signal detection.In this paper,the influence of refractive index change under different conditions such as the laser incidence angle,the angle between micro-cantilever and liquid surface is theoretically analyzed.Results show that the signal displacement(as small as 10-3) caused by the refractive index change in a traditional optical design is corresponding to the reported cantilever deflection caused by the bio-chemical reaction.That is,the detected signals of system are greatly affected by the change of liquid refractive index.The results also show that the effect of refractive index change on the signal detection can be neglected under proper parameter condition without reducing system detecting sensitivity.This theoretical result was validated by experiments.