Determination of Sulbactam Sodium in Human Plasma by HPLC-MS/MS

Ye Xiancheng,Fan Guorong,Zhu Deqiu
DOI: https://doi.org/10.3969/j.issn.1008-049X.2013.04.015
2013-01-01
Abstract:Objective: To establish an HPLC-MS/MS method for the determination of sulbactam sodium in plasma.Method: Sodium danshensu as the internal standard,the plasma samples were separated on an Agilent Zorbax SB C18 column(100 mm×3.0 mm,3.5 μm).A tandem mass spectrometer equipped with ESI was used as the detector and operated in the negative ion mode.Selected reaction monitoring(SRM) using the precursor product ion combinations of m/z 231→140 and 197→135 was used to determine sulbactam and the internal standard,respectively.Result: The established method was able to determine sulbactam sodium in human plasma over the range of 0.103~51.300 mg·L-1(r2=0.999 9)with the relative recovery from 91.62% to 102.97%.Intra-and inter-day RSDs were less than 5%.Conclusion: The method developed in the study is rapid,accurate and sensitive,and suitable for the pharmacokinetics study of sulbactam.