Fluoroimmunosensing System Using Ferulic Acid As the Substrate for the Determination of Transferrin

QY Cao,FC Gong,GL Shen,RQ Yu
DOI: https://doi.org/10.1081/al-120026420
2003-01-01
Analytical Letters
Abstract:A fluoroimmunosensing system with the use of ferulic acid as the substrate for the determination of transferrin in serum samples has been developed. It was based on paraffin-graphite-transferrin antiserum immunocomposite, which permits its surface to be easily renewed by polishing. The renewed surface serves as a platform for the competitive immunoreaction of HRP-transferrin and free transferrin (analyte) with the antibody bound to the surface. By using ferulic acid as a substrate, the amount of HRP-transferrin bound was quantified fluorimetrically, which was in turn related to free transferrin content in the samples. Ferulic acid as the substrate possesses a better stability towards H2O2 as compared to TMB. The analytical conditions, such as pH, the amount of labeled transferrin, incubation time, and temperature were studied. Under the optimized experimental conditions, a detection limit was 30 ng/mL and the linear detection range was from 39.3 to 177 ng/mL. This approach has been applied to clinical samples with satisfactory results.
What problem does this paper attempt to address?