Evaluation of visual and colorimeter shade matching

LI Zhi,GAO Ping
2008-01-01
Abstract:Objective:The aim of this study was to evaluate the accuracy of a new developed chroma meter.Methods:The chroma of shade samples was measured using shade Eye Nce with Vita classical shade guide as the standard.The measurement was considered correct if the shade value corresponded to the shade guide value of Vita shade guide,other wise it was consictered wrong.80 shade guide teeth were measured in 5 shade guide.15 observers with normal color vision were allowed to Vite-Vite test to evaluate the accuracy of Visual color match.The data were and lysed by chisquare test.Results:The rate of correct match by chroma meter is 45%,a little high than the rate of the visual color match.Conclusion:The visual mach has a higher variability among different individuals,while the chorma meter show a high consistent in color match,but the accuracy is still not satisfactory.