Experiments and Theoretical Calculation of Memory Electromagnetic Flowmeter in Simulation Well

MA Shuilong,ZHANG Yuhui,HUANG Chunhui,HU Jinhai,WANG Yanjun
DOI: https://doi.org/10.3969/j.issn.1004-1338.2013.03.003
2013-01-01
Abstract:The responses of memory electromagnetic flowmeter in water and polymer compound are studied in multiphase flow simulation well.The experiment results show the meter constant in pure water is higher than that in polymer compound,and the meter constant has no obvious change with variation of polymer compound concentration.The theoretical calculations of the responses in water and in polymer compound are made in flow field distribution and magnetic field distribution.The analysis results show that both flow velocity and relative permeability of flow media affect the responses of the flowmeter.The results of theoretical calculation are consistent well with simulation experiment results.
What problem does this paper attempt to address?