Aorta Stent Implantation Guided by Magnetic Navigation System:an in Vitro Study

WANG Su,CAI Junfeng,LUO Zhe,GU Lixu,WANG Zhe,ZHAO Qiang
DOI: https://doi.org/10.3969/j.issn.1008-6358.2012.02.010
2012-01-01
Abstract:Objective:To evaluate the feasibility,repeatability and accuracy of using magnetic navigation system to guide aorta stent implantation in virtual reality environment.Methods:Preoperative CT image information was loaded into the computer and reconstructed to a 3D anatomic structure.The next step was to complete preoperative registration with the model and to infuse the real-time ultrasound images.After that,the virtual reality environment was established.The interventional devices were transformed so that in models they can be traced accurately and located in a 3D anatomic structure by using Aurora magnetic navigation system.The aorta stent was deployed according to the sensor coil in front of the stent.The accuracy was proved by real-time ultrasound and postoperative CT.The operation time(T),registration error(E1),and the total operation error(E2) were recorded and calculated.Results: All aorta stents were successfully deployed in accurate positions.The average operation time was(38.4±3.9) min;the average registration error was(0.59±0.27) mm;the average total operation error was(1.94±1.15) mm.Registration and operation accuracy satisfied the clinical requirements.Conclusions: The experiment of in vitro models has proved that in virtual reality environment aorta stent implantation guided by magnetic navigation system is easy and repeatable.This method can be used as a new interventional tracing and positioning method.