Design on ear photoplethysmography sensor with movement interference cancellation

Dong E,Shu-ming YE,Le-chuan ZHOU
DOI: https://doi.org/10.13873/j.1000-97872012.08.004
2012-01-01
Abstract:Photoplethysmography sensor can be used to measure heart rate,oxygen saturation,respiration rate,and so on.Because of the limitation of the detecting location and the sensibility to the movement interference,it is developed for clinical application as usual,but not as home-appliance medical device.Thus,the sensor structure including LED and PD is redesigned to enhance the signal intensity,the detecting location on ear is selected by the comparison experiment,and also the 3-axis acceleration sensor is implemented to acquire the noise source for the adaptive filter which can weaken the movement interference in wallking and running state,and the heart rate can be calculated.After a sequence of activities experiment on human body,using ECG device as reference,the accuracy rate of the heart rate calculation is verified.