A High-Performance System for Weak ECG Real-Time Detection
Kun Xu,Yi Yang,Yu Li,Yahui Zhang,Limin Zhang
DOI: https://doi.org/10.3390/s24041088
IF: 3.9
2024-02-08
Sensors
Abstract:Wearable devices have been widely used for the home monitoring of physical activities and healthcare conditions, among which ambulatory electrocardiogram (ECG) stands out for the diagnostic cardiovascular information it contains. Continuous and unobtrusive sensing often requires the integration of wearable sensors to existing devices such as watches, armband, headphones, etc.; nonetheless, it is difficult to detect high-quality ECG due to the nature of low signal amplitude at these areas. In this paper, a high-performance system with multi-channel signal superposition for weak ECG real-time detection is proposed. Firstly, theoretical analysis and simulation is performed to demonstrate the effectiveness of this system design. The detection system, including electrode array, acquisition board, and the application (APP), is then developed and the electrical characteristics are measured. A common mode rejection ratio (CMRR) of up to 100 dB and input inferred voltage noise below 1 μV are realized. Finally, the technique is implemented in form of ear-worn and armband devices, achieving an SNR over 20 dB. Results are also compared with the simultaneous recording of standard lead I ECG. The correlation between the heart rates derived from experimental and standard signals is higher than 0.99, showing the feasibility of the proposed technique.
engineering, electrical & electronic,chemistry, analytical,instruments & instrumentation
What problem does this paper attempt to address?
The main problems that this paper attempts to solve are as follows: In non - invasive and continuous electrocardiogram (ECG) monitoring, it is difficult to obtain high - quality ECG signals at the wrist, arm, and ear, etc. due to the low signal amplitude and numerous interferences. Specifically:
1. **Low signal amplitude**: At the wrist, arm, and ear, etc., the amplitude of the ECG signal is very low (at the micro - volt level), which increases the difficulty of detecting high - quality ECG signals.
2. **Multiple interferences**: These parts are easily affected by multiple interferences such as power - line noise and electromyogram (EMG), resulting in a decline in signal quality.
To solve these problems, the author proposes a high - performance multi - channel signal superposition system for the real - time detection of weak ECG signals. Through multi - channel signal superposition, interference can be effectively suppressed without relying on complex signal processing, thereby improving the signal - to - noise ratio (SNR). This system includes an electrode array, a collection board, and an application (APP), and its feasibility has been verified through experiments. For example, in experiments carried out on the ear and upper arm, this system has achieved a signal - to - noise ratio of more than 20 dB, and the correlation with the standard lead I ECG recording is higher than 0.99.
In summary, this paper aims to develop a system that can achieve high - precision, real - time ECG monitoring at the wrist, arm, and ear, etc., to meet the needs of home medical monitoring.