To Study the Value of Computer-Aided Detection in CT Colonography in Colonic Polyp Diagnosis

李野,程凯亮,张春雨,张惠茅
DOI: https://doi.org/10.3969/j.issn.1007-4287.2010.01.029
2010-01-01
Abstract:Objective To explore the value of computer-aided detection in multi-layered CT colonography in colon polyp diagnosis.Methods CTC (with patients in both supine and prone positions) was performed with GE 64-MD CT scanner in 23 patients highly suspected with colon polps.Sixty colon polyps were confirmed by colonoscope,of them ten was smaller than or equal to 5 mm,twenty-one was 6-9 mm and twenty-nine was greater than or equal to 10 mm in diameter.Three readers blinded to the result of fibrocolonoscope read the result without and then with computer-aided detection.The diagnostic result would be confirmed at least two of the three readers gave the same diagnosis The sensitivity of detection polyps without and with computer aided detection was assessed.Results With CAD as a second reader,the sensitivity of detecting polyps smaller than or equal to 5 mm in diameter increased from 40% to 60%,P0.05;it increased from 67% to 86% for 21 colon polyps between 6-9 mm in diameter,P0.05;sensitivity of detecting bigger than or equal to 10mm polyps increased from 90% to 93%,P0.05.Conclusion Computer aided detection is helpful in the multi-layered CT colonography for detecting polyps smaller than 10mm in diameter.