Abstract:The semi-volatilization organic compounds(SVOCs) in water were determined by GC-MS method combined with solid phase extraction(SPE).SVOCs in water samples were extracted and concentrated with SPE column of Waters Porapak Rdx.The DB-1701 capillary column(30 m×0.25 mm×0.25 μm)was used to separate SVOCs,and then each compound in SVOCs was determined by MS method in which an EI ion source was used and selected ion monitoring mode(SIM) was applied.The results showed that SVOCs in water samples were effectively enriched and separated by SPE/GC method.A good linear relationship(R=0.990) was obtained between the detector respond values and concentration of SVOCs in the range of 0.05-0.40 mg/L,and the detection limits(3 S/N) were in the range of 0.13-15.87 μg/L.Recovery and precision tests were performed by standard addition method using a blank water sample,and gave a recovery of 70%-112% with RSD(n=6) of 2.67%-17.17% for each semi-volatilization organic compound.The method is suitable for simultaneous determination of dozens of SVOCs contained in environmental quality standards for surface water and standards for drinking water quality.The results could be used as references for water quality monitoring and further studies.