Proper extrinsic positive end-expiratory pressure during mechanical ventilation in patients with chronic obstructive pulmonary disease

SHENG Hua,ZHU Huili,LI Xiangyang,FU Ligang,CHEN Lin
2009-01-01
Abstract:Objective To discuss the proper range of extrinsic positive end-expiratory pressure (PEEPe) during mechanical ventilation in patients with chronic obstructive pulmonary disease (COPD). Methods Thirty-six ventilated patients with confirmed COPD were included in the present study. Dynamic intrinsic positive end-expiratory pressure (PEEPi),work of breathing patient (WOBp),work of breathing ventilation (WOBv),difference of esophagus pressure (dPes),tidal volume (Vt),peak pressure (Ppeak),plateau pressure (Pplat),and respiratory system compliance (Crs) were measured by esophagus balloon pressure sensor and flow sensor. Heart rate (HR),systolic blood pressure(SBP),diastolic blood pressure(DBP),cardiac output(CO),and cardiac index (CI) were measured by GE Solar 8000 monitor. All above measurements were repeated with PEEPe set at 0,40%,50%,60%,70%,80%,90% and 100% of PEEPi. Results WOBp decreased as PEEPe increased,and the difference between PEEPe at 60% of PEEPi and baseline was significantly different (all P<0.05). The dPes decreased as PEEPe increased,and the difference between PEEPe at 70% of PEEPi and baseline was significant (all P<0.05). Ppeak and Pplat increased with the increase of PEEPe,and significant difference was found between PEEPe at 60% of PEEPi and baseline (all P<0.05); besides,the difference between PEEPe at 60% of PEEPi and PEEPe at 100% of PEEPi was also significant (P<0.05). When PEEPe increased,CO and CI decreased continuously,and there was significant difference between PEEPe at 90% of PEEPi and baseline (both P<0.05). Conclusion PEEPe at 60%-80% of PEEPi can reduce WOBp and improve Crs. PEEPe at this level has little inhibition on circulation and is the optimal PEEPe for patients with COPD.
What problem does this paper attempt to address?