Measurement of Glucose Concentration in Blood Plasma Based on a Wireless Magnetoelastic Biosensor

Pengfei Pang,Wenyue Yang,Sijing Huang,Qingyun Cai,Shouzhuo Yao
DOI: https://doi.org/10.1080/00032710701242097
2007-01-01
Analytical Letters
Abstract:A wireless magnetoelastic glucose biosensor in blood plasma is described, based on using a mass sensitive magnetoelastic sensor as transducer. The glucose biosensor was fabricated by coating the ribbon-like, magnetoelastic sensor with a pH sensitive polymer and a biolayer of glucose oxidase (GOx) and catalase. The pH response polymer swells or shrinks, thereby changing sensor mass loading, respectively, in response to increase or decrease of pH values. The GOx-catalyzed oxidation of the glucose in blood plasma produces gluconic acid, resulting in the pH sensitive polymer shrinking, which in turn decreases the sensor mass loading. The results show that the proposed magnetoelastic glucose biosensor can be successfully applied to determine the concentration of glucose in blood plasma. At glucose concentration range of 2.5-20.0 mmol/l, the biosensor responses are reversible and linear, with a detection limit of 1.2 mmol/l. Since no physical connections between the sensor and the monitoring instruments are required, this proposed biosensor can potentially be applied to in vivo and in situ measurement of glucose concentration in physiological fluids.
What problem does this paper attempt to address?