A new piezoelectric quartz crystal immunosensor for hCG detection

Fu Ting,Wang Hua,Shen Guo-li,Yu Ru-qin
DOI: https://doi.org/10.3969/j.issn.1008-2298.2006.02.003
2006-01-01
Chemical Sensors
Abstract:A new piezoelectric quartz crystal immunosensor for hCG detection was constructed. The anti-hCG antibody was deposited on the gold electrode's surface of 9 MHz quartz AT-cut crystal by self-assembled cysteamine technique and carbonylic carbon nanotube. Compared with conventional antibody immobilization methods, the sensitivity of the sensor coated by cysteamine and carbonylic carbon nanotube is improved than those by cysteamine only. Under the optimized experimental conditions, the results showed that the low detection limit was 0.01 mIU/mL, and the linear range was 0.06~149.12 mIU/mL.