GC Determination of Alantolactone and Isoalantolactone in Zangmuxiang

WANG Yan,ZHANG Ya-hui,GEsang Suo-lang,SUN Wen-ji
2009-01-01
Abstract:Objective:A capillary GC method was established for the determination of alantolactone and isoalantolactone in Zangmuxiang.Method:The analytical column was SE-30 capillary column(30 m×0.32 mm×0.25 μm)with an FID detector.The injector temperature was 240 ℃.The detector temperature was set at 260 ℃.The column temperature was programmed from 50 ℃ at a rate of 20 ℃·min-1 to 170 ℃(holding 28 min),then raised to 250 ℃ at 25 ℃·min-1.The final temperature was maintained for 1 min.Results:The linear ranges of alantolactone and isoalantolactone were 0.6-4.2 mg·mL-1 (r=0.9999),which showed good linearity.Conclution:The methed is simple,specificity,accurate and reliable,and can be used for quality control of Zangmuxiang.