INFLUENCE OF TITANIUM ALLOY CD-CERVICAL ON MAGNETIC RESONANCE IMAGE OF PATIENTS WITH POSTERIOR CERVICAL SPINAL FUSION

Gao Yujun,He Shisheng,Hou Tiesheng
DOI: https://doi.org/10.3969/j.issn.1007-9564.2008.11.012
2008-01-01
Abstract:Objective To discuss influence of titanium alloy CD-cervical internal fixation system on magnetic resonance image(MRI) of cervical spinal cord. Methods 8 cases who had operated in our hospital with titanium alloy CD-cervical internal fixation system performed MRI examimation. The MRI system was 1.0 or 1.5 T scanner, patients all performed sagital T1, T2 and axial T1-weighted image in spin echo mode, then the artifacts and its influence on the structure of spine canal were analyzed. Results The implant of titanium exhibit low intensity and the margin of the implant exhibit high intensity in T1 and T2-weighted image. The laminar hooks of CD-cervical exhibit significant artifacts which deform the structures of spine canal, above the laminar hook, CD-cervical exhibit small artifacts, the structures of spine canal were clear. Conclusion Titanium alloy CD-cervical internal fixation system plates exhibit significant artifacts on MRI, it is a disadvantage to postoperative follow-up and evaluation.
What problem does this paper attempt to address?