Simultaneous Determination of 42 Amino Acids in Human Body by Stable Isotope Isobaric Tage for Relative and Absolute Quantitation Labeled-LC-MS/MS with Typical Example Cases

Li Peng-Fei,Wang Yan,Tao Bei-Bei,Wang Jing,Luo Jing,Zhang Xu-De,An Zhuo-Ling,Liu Li-Hong
DOI: https://doi.org/10.3724/sp.j.1096.2012.10917
2012-01-01
Abstract:To develop a LC-MS/MS method for the simultaneous determination of 42 amino acids in human body by stable isotope isobaric tage for relative and absolute quantitation(iTRAQ) labeled.After preciptitated by sulfosalicylic acid and labeled by iTRAQ reagent 115,the biological samples were added iTRAQ reagent 114-labeled amino acid as isotope internal standard,and separated on an AAA-C18 column using water and acetonitrile(both containning 0.01% heptafluorobutyric acid and 0.1% formic acid) as mobile phase by gradient elution.Detection was carried out by multiple reaction monitoring(MRM) on a 3200QTRAP LC-MS/MS system.Quantitation was completed by isotope internal standard which also could remove the systematic error.42 amino acids and isomers were chromatographic baseline separated.The assay was successful applied to clinical diagnose and treat.The method is a rapid,sensitive,selective and high-flux for the determination of amino acids metabolism disease and nutritional evaluation.
What problem does this paper attempt to address?