Determination of S-adenocyl-L-methionine by High Performance Liquid Chromatography

Liu Zhuang,Jiao Yuhai,Liu Yan,Liu Xuesong
DOI: https://doi.org/10.3969/j.issn.1008-2565.2009.04.005
2009-01-01
Abstract:In this work,quantitative analysis of S-adenocyl-L-methionine was proposed.The high performance liquid chromatographic behavior of toluenesulfonic acid disulfuric acid S-adenocyl -L-methionine(SAM) was researched in the high performance capillary electrophoresis liquid chromatography.This methods mainly uses C_(18) column by using ammonium formate-watemethanol -sodium octane sulfonate ion pair mixture solution as mobile phase with 1.0 mL/min flow ratio and UV detection at 252 nm.The results showed that under the optimum conditions the peak intensity was linearly related to the concentration of SAM in the range from 0.005 mg/mL to 0.25 mg/mL with the detection limit of 0.001 mg/mL.The samples recoveries ranged from 92.6%~101.1%with the relative standard deviations less than 0.3381%in the linear range.Quantitative analysis of S-adenocyl-L-methionine was studied and the good results were obtained.