Application of continuous noninvasive hemoglobin monitoring in patients undergoing liver transplantation

ZHU Hui,CAO Qiang,ZHANG Yan,SU Diansan,CHEN Yongming,WANG Xiangrui
2013-01-01
Abstract:Objective To evaluate the value of continuous noninvasive hemoglobin (SpHb) measured by Masimo Radical-7 Pulse CO-Oximeter in patients undergoing liver transplantation. Methods Twenty-four adult patients undergoing orthotopic liver transplantation were enrolled in this study. SpHb was monitored continually and recorded before induction of anesthesia, 5 minutes after induction of anesthesia, 5 minutes after skin incision, 5 minutes before anhepatic phase, at anhepatic phase, 5, 30 and 90 minutes in neohepatic phase, and at the end of surgery (the data at the last 8 time points was regarded as intraoperative SpHb). The arterial blood samples were collected and invasive hemoglobin (THb) levels were measured at above mentioned time points. THb levels were compared with the corresponding SpHb. Results SpHb had a linear correlation with THb before induction (r=0.81, P<0.001) and during the surgery (r=0.75, P<0.001). The regression equations were y=1.66+0.785x and y=0.589+0.766x, respectively. The mean difference between SpHb and THb was 9 g/L before induction and 20 g/L during the surgery by Bland-Altman consistency analysis with 95% CI being -26-44 g/L and -15-55 g/L, respectively. Conclusion There is a good consistency between SpHb and THb in the patients undergoing liver transplantation. But the accuracy of intraoperative SpHb is lower than that of preoperative one.
What problem does this paper attempt to address?