GC/MS Analysis of Malathion Technical and its Main Impurities

Gu Kai,Li Zenghe
DOI: https://doi.org/10.3969/j.issn.1002-5480.2012.02.011
2012-01-01
Abstract:To establish a method for the determination of malathion technical by GC/MS(EI).The GC/MS system consisted of HP-5MS column and electron impact ion source.The standard curve of malathion showed good linearity over the range of 1.5~10mg/L.The standard deviation was 0.93,the variation coefficient was 1.0%,and the average recovery of the method is 99.13%.The method is simple,accurate and reproducible.