Biocompatible study on polyimide film electrode with optic nerve glial cells of rat

Chen Wei,Zheng Hua,Liu Peipei,She Zhenjue,Lu Yiliang,Ren Qiushi,Zhou Guomin
2009-01-01
Abstract:To evaluate the adhesion,biocompatibility between glial cell and polyimide(PI) film electrode and the best parameters of electrical pulse in vitro,providing scientific basis for the implant of artificial vision substrate material,the research began with the culturing glial cells of rats in vitro and performed manifold biological experimental methods including cytotoxicity test and adhesion test to evaluate the biocompatibility of substrate material systemically.Established the electrical stimulation model of glial cells of rats in vitro and observed the biology characteristic's of pulse current to the glial cell.The results showed that reaction scales of polyimide film electrode cytotoxicity were 0 to 1 level,no significant cytotoxicity.The adhesion was fine and no obvious changes in cultured cells attached to the surface of film electrode.The growth and proliferation of the cultured cells were not significantly inhibited with the substrate materials.Under the scanning electron microscope,the glial cells were observed closely attached to the surface of film electrode,the good spreading and the connecting into the film.At a regular pulse width and amplitude as well as the stimulation time,the frequency was lower,and the impact on cell was smaller.The present results suggest that the polyimide thin film electrode was well adhesion and biocompatibility with the glial cell of rat,and might form an effective electrical pulse stimulation which is appropriate for optic nerve implant.