Flow chemiluminescence sensor for determination of clenbuterol based on molecularly imprinted polymer

Houjiang Zhou,Zhujun Zhang,Deyong He,Yufei Hu,Ying Huang,Dinglong Chen
DOI: https://doi.org/10.1016/j.aca.2004.06.068
IF: 6.911
2004-01-01
Analytica Chimica Acta
Abstract:In this paper, a novel flow chemiluminescence (CL) clenbuterol sensor based on molecularly imprinted polymer (MIP) on line enrichment nanogram clenbuterol and chemiluminescence reaction of potassium permanganate and formaldehyde in the polyphosphate enhanced by clenbuterol. Clenbuterol in the urine was selectively adsorbed on the clenbuterol-imprinted polymer, which was packed into the flow cell. The formaldehyde and the polyphosphate with potassium permanganate flowed through the flow cell and reacted with the on line adsorbed clenbuterol and produced strong CL. The results show that the sensor was reversible. The CL intensity was linear with clenbuterol concentration from 1.0 × 10−9g/mL to 5.0 × 10−8g/mL. The detection limit was 3.0 × 10−10g/mL. The R.S.D. for ng/mL clenbuterol was less than 5% (n = 3). The present method offered a high selectivity and sensitivity that made the quantitative analysis of trace clenbuterol (ng/mL) in the animal urine sample.