Detection of complement C1-inhibitor with a piezoelectric immunosensor

Li Liu,Jiming Hu,Leilei Wang,Lijia Liu,Xiaodong Zhou
DOI: https://doi.org/10.1007/s002160000657
2001-01-01
Abstract:A novel piezoelectric immunosensor has been developed for the detection of human complement C1-inhibitor. Anti-C1-inhibitor antibody was immobilized onto the gold electrodes of a 9 MHz AT-cut piezoelectric crystal. Coating the crystal with polyethyleneimine adhesion, followed by a glutaraldehyde cross-linking method to immobilize antibody showed better results than the physical adsorption method with respect to sensitivity and reproducibility. Under the optimized experimental conditions, the sensor showed good response to the C1-inhibitor in the range from 2.0 × 10 –8 to 1.2 × 10 –6 g. Other proteins in human serum did not remarkably interfere with the detection. The crystals could be regenerated 5 times, when bound materials on the crystal surface were eluted by strong acid and strong alkali solution and subsequently cleaned in an ultrasonic cleaner.
What problem does this paper attempt to address?