Research on Near-Infrared Spectrum Detecting Circuit with LIA

JIA Zhen-an,WANG Pei-pei,QIAO Xue-guang,ZHOU Hong,WANG Wei
DOI: https://doi.org/10.3969/j.issn.1002-1841.2012.04.037
2012-01-01
Abstract:Lock-in amplify(LIA) is one of the important techniques for weak signal detection,so it is widely used in near-infrared spectral measurements.Based on the principle of spectrum absorption,this paper used phase shifting that by adjusting the size of the bias voltage controls the phase difference between the detected signal and reference signal,and the use of balance modemAD630 which achieves phase-sensitive detection were effective on restraining noise to improve the performance of the system.The results indicate that the output of the detective circuit is linear with the gas volume fraction,its linearity of fitting coefficients is 0.998 5,and the sensitivity is 0.045 7 V/%.