Determination of Tyrosine by the Combination of Ion chromatography,online Post-column Electrochemical Derivatization and Fluorescence spectrometry

SHI Weiwei,LI Gang,ZHENG Yimei,LV Weide,WU Shuchao,ZHU Yan
2013-01-01
Abstract:Using a self-constructed electrolytic cell for electrochemical derivatization treatment,a novel method based on ion chromatography combined with online post-column electrochemical derivatization and fluorescence detection was developed for the determination of tyrosine.Following the separation on the anion exchange column with alkaline eluent,tyrosine was introduced to the anode chamber of the electrolytic cell for electro-oxidization,of which oxidization product has strong fluorescence yield thus can be easily detected by fluorescence detector.Ionization reagent was used both as the eluent for separation and the supporting electrolyte for electrochemical reaction,which guarantee the good compatibility between the ion chromatography and post-column electrochemical derivatization.The optimal experimental conditions are as follows:eluent of NaOH(10mmol/L)+acetonitrile(1+9),flow rate of 1.0mL/min,electro-oxidation potential of 1.0 V,excitation and emission wavelength of 320 nm and 420 nm,respectively.Under these optimal experimental conditions,the linear range of tyrosine was from 0.01to10mg/L and the detection limit was 1.2μg/L with the signal-to-noise ratio of 3.After seven replicate injections of 50μg/L tyrosine,the relative standard deviation of the chromatogram peak area was 2.5%.The method has advantages of rapidity,high sensitivity and good selectivity and was successfully applied for the determination of tyrosine in drink.