Determination of fluvastatin concentration in human plasma by high performance liquid chromatography with fluorescence detection

PEI Qi,LI Zuo-jun,YANG Guo-ping,ZHOU Yu-lu,YI Dan,HUANG Qi,WU Cui-fang,LIU Shi-kun
DOI: https://doi.org/10.3969/j.issn.1007-7669.2006.10.009
2006-01-01
Abstract:AIM: To establish a simple and rapid method for determination of fluvastatin concentration in human plasma by HPLC with fluorescence detection. METHODS: Chromatographic column was Kromasil C18 (4.6 mm×150 mm, 5μm) with 40℃of column temperature. The mobile phase consisted of 13 mmol·L-1 tetrabutylammonium chloride-acetonitrile (51 : 49, V/V) with a flow rate of 1.0 mL·min-1 . A fluorescence detector was set atλex305 nm andλem405 nm. RESULTS: The lower detectable limit of fluvastatin was 1μg·L-1. The calibration curve was in good linearity within the range of 1 - 1 000μg·L-1 (r = 0.999 4) . The intra-day and inter-day RSD were less than 10 %. CONCLUSION: The method of HPLC with fluorescence detection is sensitive, rapid, and suitable for the study of pharmacokinetics of fluvastatin.