THE COMPARISON OF TWO METHODS FOR CHOLESTEROL DETERMINATION IN EEL BY HPLCAND CHOLESTEROL KIT

TAN Guan-yue,FU Hong,YANG Lin
DOI: https://doi.org/10.3969/j.issn.1005-6521.2008.04.046
2008-01-01
Abstract:Cholesterol was quantified by reversed phase liquid chromatography on a C18(4.6 mm×250 mm, 5 μm)using a mobile phase of acetonitrile and 2-p ropanol (60∶40) with a flow rate of 1.0mL/min. A linear correlation was observed between 6 mg/L-600 mg/L of cholesterol with a good correlation coefficient (r=0.999 47) between the content of cholesterol and peak area. The determination limit was 1mg/L. The mean recoveries were 98.75 %,100.7 %,102.75 %for the low, middle and high levels of addition, respectively. The coefficient of variation was 0.4 %. There was obvious difference of cholesterol of eel quantified by HPLC and by cholesterol Kit. The cholesterol content quantified by cholesterol kit was 67.7 % higher than by HPLC. This paper also discussed the reason of high cholesterol of eel quantified by cholesterol kit.