Determination of 8-oxo-deoxyguanosine in human urine by HPLC-ECD method

Wang Tian-cheng,Jia Guang,Wang Xiang,Yan Lei,Shen Hui-qi
DOI: https://doi.org/10.3969/j.issn.1004-8685.2006.04.009
2006-01-01
Abstract:Objective:To establish a HPLC-ECD method to measure 8-oxo-deoxyguanosine(8-OHdG) in human urine.Methods:The urinary sample was pretreated with a solid phase extraction(SPE) column.Then the chromatography was performed by an ODS C_(18) column with ECD detector and mobile phase of acetic acid buffer which contained 8% methanol.Results:The linear range was 25~400 μg/L,r=0.9892,the recovery rate was 91.2%~108.8%,the intra-and inter-day precisions were 3.9%~7.5% and 4.3%~8.6% respectively,the detection limit was 2 μg/L.Conclusion:It is a better method for the determination of 8-OHdG in human urine.
What problem does this paper attempt to address?