A Sensitive Biosensor Based on Os-complex Mediator and Glucose Oxidase for Low Concentration Glucose Determination

Liying Jiang,Hongmin Liu,Jian Liu,Qingde Yang,Xinxia Cai
DOI: https://doi.org/10.1016/j.jelechem.2008.03.018
IF: 4.598
2008-01-01
Journal of Electroanalytical Chemistry
Abstract:The determination of low concentration glucose is important with the development of noninvasive glucose monitoring technology. This paper describes a sensitive biosensor formed by the immobilization of Os-complex mediator and glucose oxidase on the electrode surface. Glucose biosensor exhibited analytic sensitivity of 28.24nAμM−1cm−2 at a low applied potential (0mV vs. Ag∣AgCl). Response to glucose proves to be linear with the correlation coefficient up to 0.999 in the relevant concentration range of 0–700μM. Relative standard deviations were 4.75% and 4.62% for within-run precision (n=10) and between-run precision (n=11), respectively. The limit of detection for glucose was 0.3μM. The biosensor exhibited sensitivity of 4.67nAmM−1cm−2 for subcutaneous glucose. The biosensor also indicates good stability of more than 90% of its original activity for about 15 months stored at 4°C. The results show the biosensor is highly sensitive for the low concentration glucose, especially for extracted subcutaneous glucose.
What problem does this paper attempt to address?