Thoracolumbar Pedicle Screw Fixation with C-arm Navigation

沈点红,孔荣,黄炎,方诗元,禹德万,李守民,张光平,刘兵
DOI: https://doi.org/10.3969/j.issn.1008-0287.2009.02.004
2009-01-01
Journal of Clinical Orthopaedics
Abstract:Objective To assess the accuracy of computer navigation technology used in thoracolumbar pedicle screw fixation.Methods 15 cases with thoracolumbar lesions underwent orthopaedic operations.72 pedicle screws were inserted with the computer navigation.After the operation,the position of the pedicle screw was evaluated with CT and X-ray using the Richter grading system.Results All cases succeeded with assistance of computer navigation.15 cases were followed up for 6 to 24 months.No pedicle screw loosening or breakage was found.According to the location of the screw in X-ray and CT results,there was no vessel,nerve or spinal cord complication.According to Richter grading system,the results were excellent in 68 cases(94.4%) and good in 4(5.6%).It obviously reduced the radiation exposure to C-arm fluroscopy and the time that spend on surgery.It obviously improved the success rate of surgery.Conclusions C-arm navigation can obviously reduce error in thoracolumbar pedicle screw fixation and improve the accuracy of it.