Hyperthermia by a Nitinol Stent in an Alternating Magnetic Field: Safety and Feasibility in Rabbit Esophageal Cancer

Jumei Zhou,Ning Li,Qisheng Xia,Lehui Du,Xiaowen Wang,Linyun Zhao,Xufei Wang,Jintian Tang
DOI: https://doi.org/10.1016/j.pnsc.2009.07.013
2009-01-01
Abstract:Hyperthermia by heating in-stent is a potentially new method to treat esophageal cancer. This study assessed the safety of heating the rabbit esophagus with a nitinol stent in an alternating magnetic field (AMF) and investigated whether this method offers a therapeutic option for esophageal cancer. Nitinol stents were placed in the cervical esophagi of healthy rabbits, which were heated in an AMF for different times at 43, 46 and 50°C. The esophagi were histologically examined after a week to observe whether there was transmural necroses. Then esophageal cancer of rabbit placed nitinol stents were heated in an AMF at 46°C for 10min. Stents were heated to a target temperature within 5min. The highest tolerated temperature and time for the healthy rabbit esophagus was at 46°C for 10min. Tumor growth was delayed by heating and it was statistically significant. Heating the rabbit esophageal wall at 46°C for 10min proves to be safe and effective in delaying tumor growth.
What problem does this paper attempt to address?