The Effect of the Closely-Spaced Working and Auxiliary Electrodes on the Performance of Electrochemical Oxygen Sensor

Jian Wu,Zunzhong Ye,Yibin Ying,Feng Ji
DOI: https://doi.org/10.1002/elan.200703950
IF: 3.077
2007-01-01
Electroanalysis
Abstract:In this paper the influence of the electrochemical reaction at the auxiliary electrode of oxygen microsensors on the sensor performance was investigated. When the auxiliary electrode is closely spaced to the working electrode, the redox cycling of O-2/H2O takes place in an electrochemical oxygen sensor. This cycling alters the oxygen distribution around the working electrode and therefore affects the measured cathodic current passing through the working electrode. Calibrations have to be taken out to determine the real O-2/H2O cycling effects. Furthermore, this redox cycling also provides the possibility to enhance the sensitivity of the electrochemical oxygen microsensor. Experimental results indicate that the sensitivity of the oxygen sensor is enhanced 3.0 times with the on-chip 10 mu m spaced interdigitated auxiliary electrode.
What problem does this paper attempt to address?