Determination of Nitrite and Nitrate-Nitrogen in Seawater by Flow Injection Analysis Based on Quantificational Pipe Clamp Solenoid Valves

Han Bin,Gao Lei,Zheng Li,Song Zhuan-Ling,Jiang Feng-Hua,Zang Jia-Ye,Wang Xiao-Ru
DOI: https://doi.org/10.3724/sp.j.1096.2010.01832
2010-01-01
Abstract:Using three pipe clamp solenoid valves to replace the traditional six-port valve for sample quota,a set of multichannel flow injection analyzer was designed for the determination of nitrate and nitrite in seawaters.The construction of apparatus is simple and has potential to be used for analysis of nitrite and nitrate.The sample throughput of nitrite and nitrate was 28 samples/h.The liner range of nitrite and nitrate was 25.0-1000 μg/L and 50.0-1000 μg/L(R≥0.999).The detection limits of nitrite and nitrate were 1.81 μg/L and 4.82 μg/L.The recovery of nitrate and nitrite was 94.5%-105.3%.The relative standard deviations for 200 μg/L of nitrate and nitrite was 1.3%-4.5%(n = 6).The results showed that there are not significance differences between this method and national standard method in the t-test analysis.The method is suitable for fast analysis of nitrite and nitrate in seawater.