Detection of Malathion in Seawater Using a Flow Injection Type Biosensor Based on the Immobilized Acetylcholinesterase

孟范平,何东海,朱小山,杨正先,马冬冬
IF: 1.193
2005-01-01
Chinese Journal of Analytical Chemistry
Abstract:With the help of a flow injection analysis (FIA) system, the possibility to determine malathion in seawater continuously by the biosensor based on the immobilized acetylcholinesterase (AChE) was studied. In this system, the carrier was seawater without malathion. The feasible work conditions for the biosensor were as follows: the flow rate was 0.39 mL/min for both the carrier and samples; the inhibition time from the samples lasted for 20 minutes; the concentration and injection volume of the substrate, acetylthiocholine iodide, were 41.6 mmol/L and 40 μL, respectively. The sensitivity of the biosensor had a remarkable rise when the samples were oxidized previously by hypochlorous sodium. The detection limits of the biosensor for malathion in seawater were 1.3 μg/L and 0.05 μg/L before and after pre-oxidation respectively. In addition, for the samples with malathion from 0.1 μg/L to 10 μg/L, a good linearity (r=0.991) between the response of biosensor and the concentration of malathion could be obtained. After a sample containing malathion less than 100 μg/L was measured, the inhibited activity of immobilized AChE could be reactivated completely by transporting pyridine-2-aldoxime of 0.5 mmol/L into the biosensor for 15 minutes. These showed that the biosensor was siutable for the sensitive, accurate and continuous detection of organophosphorus pesticides in seawater.