An Experimental Model of Membraneous Oxygenator

ZHAO Hong-ming,DONG Gang,WU Heng,LIU Guo-qing
DOI: https://doi.org/10.3969/j.issn.1009-7090.2006.01.004
2006-01-01
Abstract:Objective To test the oxygenation property of membraneous oxygenator by means of the model established simulating invo and ex vivo gas exchange in circulation.Methods The model established respiratory circulation gas exchange was connected to the testing membraneous oxygenator with polyvinyl chloride tube.Thus the close loop of circulation was completed.The fresh bovine blood added anticoagulant was adopted as test medium.The blood samples were taken from the inlet and outlet of tested oxygenator in 3,5,60,120,240 and 360 minutes for gas analysis and blood routine examination.Then compared the oxygenation property of testing Xi'an Xijian Ⅰ oxygenator and Affinity(USA) oxygenator with the model.Results It demonstrated by the model there was good oxygenation property of the testing home made oxygenator,the SaO2,PaO2 and PaCO2 were 95 %,13.3 kPa(100 mmHg) and 5.33 kPa(40 mmHg) respectively after circulating 360 minutes.Conclusion It is demonstrated that the construction of the model is satisfying the fundamental principle of extracorporal circulation and it is reliable and applicable for the study on properties of membraneous oxygenator and fundamental principle of extracorporal circulation.