Accuracy in Evaluation of Strain and Strain Rate by Doppler Tissue Imaging in Regional Myocardial Function Detection Compared with Velocity Vector Imaging

陈明,刘怡,苗爱雨,张宇辉,刘盛林,郑建华,张波
DOI: https://doi.org/10.3969/j.issn.1000-3614.2009.04.011
2009-01-01
Abstract:Objective:To evaluate the accuracy of strain and strain rate(SR)of Doppler tissue imaging(DTI)in regional myocardial function detection compared with velocity vector imaging(VVI). Methods:Twenty patients with left ventricular abnormal wall motion(AWM)and 20 healthy subjects were studied.Strain and SR obtained by DTI were compared with that by VVI. Results:Statistic difference in strain and SR by VVI was observed from both normal and abnormal segments(P0.05~0.001).Compared with VVI,the strain by DTI had no significant difference between normal and abnormal segments(10.63~18.81%vs.10.58~15.45%in average;P0.05)except for the basal lateral segments(7.66±8.06%vs.4.95±3.29%;P =0.01).SR by DTI had no significant difference between normal and abnormal segments(1.17~1.94 s~(-1)vs.0.73~1.51s~(-1) in average;P0.05~0.01)except for the basal lateral segments(7.66±8.06%vs.4.95±3.29%;P=0.01).All data in SR by DTI was decreased in abnormal segments(P0.05~0.01)in terms of AWM,in spite of the segment of apex and the basal lateral segments(l.19±1.83 s~(-1)vs.0.93±0.70 s~(-1);P0.05).There was poor consistency in the measurement of strain and SR between VVI and DTI by the Bland-Altman analysis. Conclusion:Strain and SR detected by DTI seemed to have inherent limitation compared with that by VVI.