An Amperometric Immunosensor for Microcystin-(Leucine-Arginine) Based on Screen-Printed Carbon Electrode

CHEN Xiang-Qiang,HE Miao,SHI Han-Chang,CAI Qiang
DOI: https://doi.org/10.3724/SP.J.1096.2011.00443
IF: 1.193
2011-01-01
Chinese Journal of Analytical Chemistry
Abstract:An amperometric immunosensor for the detection of microcystin-(leucine-arginine) (MCLR) was prepared by immobilizing an MCLR-ovalbumin (MCLR-OVA) conjugate to an osmium bipyridyl complex-attached poly(4-vinylpyridine) membrane modified onto a screen-printed carbon electrode. The immunosensor Was based on the indirect competitive immunoassay format with the labeled agent of horseradish peroxidase-conjugated goat anti-mouse immunoglobulin-G antibody. The current response decreased proportionally with the concentration of MCLR in the range of 0.43-10.72 mu g/L with a detection limit of 0. 17 mu g/L. The mean recoveries of standard additions in three real water samples were in the range of 83% to 121%, and the RSD were in the range of 3. 6% to 7.6%. Thus, the proposed method is a promising technique for the detection of MCLR in environmental monitoring.
What problem does this paper attempt to address?