Application of lens density measurement in grading diagnosis of age related cataract

裴雪婷,鲍永珍,黎晓新
DOI: https://doi.org/10.3760/cma.j.issn.2095-0160.2009.01.012
2009-01-01
Abstract:Objective The examination methods for cataract assessment can be divided into two types,subjective and objective.This study was to assess the parameters of lens density measured by Pentacam Scheimpflug system in grading diagnosis of age-related cataract.Methods 351 cases(480 eyes) with unique type of age-related cataract were recruited for this study and were divided into nuclear opacity NO1-5 and nuclear colour NC1-5,cortical cataract C1-4 posterior subcapsular cataract P1-4 based on LOCS Ⅲ grading system and 30 eyes for each.The lens density value measured by Pentacam Scheimpflug system was evaluated and the correlation between the lens density value with LOCS Ⅲ grading scores or LogMAR visual acuity was analyzed.Results There was a linear positive correlation between LOCS Ⅲ nuclear score and lens density value.The stronger positive correlation existed between LogMAR visual acuity and nuclear lens density value(r=0.867,P<0.01) than that of LogMAR visual acuity with nuclear opacity score(r=0.741,P<0.01) and nuclear colour score(r=0.653,P<0.01).Lens density value showed a weak correlation with LOCS Ⅲ cortical score(r=0.634,P=0.01) or ACD(r=0.453,P=0.02) or ACV(r=-0.380,P=0.04).No statistical significant correlation was found between lens density value and LogMAR visual acuity in cortical cataract and posterior subcapsular cataract,between LogMAR visual acuity and cortical score or between LogMar visual acuity and posterior subcapsular cataract score.Conclusion Lens density,as a quantitative and objective parameter,can offer the evaluation of degree of nuclear opacity and associated visual impairment caused by nuclear cataract.The combination of lens density value with ACD is helpful for assessing the progression of cortical cataract.
What problem does this paper attempt to address?