Determination of Avermectin in Wastewater by High Performance Liquid Chromatography with Solid-phase Extraction

LI Hong-ying,MA Peng-fei
DOI: https://doi.org/10.16241/j.cnki.1001-5914.2011.08.004
2011-01-01
Abstract:Objective To establish a high performance liquid chromatography with solid-phase extraction method for the determination of avermectin in wastewater.Methods The water sample was separated and purified after passing through Sep-Pak cartridges.Avermectin in extraction column was eluted by 95% ethanol and its content was determined by HPLC,with methanol+ water(95+5) as mobile phase and UV-detection wavelength at 244 nm.Results Linear relationships between peak area and concentration of avermectin was in the range within 1-8 μg/L(R2=0.997 7) and detection limit for avermectin was 19.3 μg/L.The precision and recovery rate were in the ranges of 4.8%-5.3% and 92.8%-104.0% respectively.Conclusion This method is simple,fast,accurate and applicable to the determination of avermectin in wastewater.
What problem does this paper attempt to address?