Finite-Element Analysis of Injectable Calcium Sulfate Bone Cement Augmentation with Dynamic Hip Screw System for the Treatment of Osteoporotic Intertrochanteric Fractures

YU Xiao-wei,WANG Nan,TANG Ting-ting
DOI: https://doi.org/10.3321/j.issn:1006-2467.2009.11.029
2009-01-01
Abstract:The mechanical performance of the dynamic hip screw (DHS) system augmented with calcium sulfate bone cement (CSC) for the fixation of stable osteoporotic intertrochanteric fracture was evaluated by means of finite element analysis (FEA). Finite element model was used to simulate the stable osteoporotic intertrochanteric fracture with DHS fixation or DHS fixation plus CSC augmentation and the stress distribution was analyzed. With cement augmentation, 35% reduction of the maximum stress was found in the femur head, suggesting reduced possibility of screw cut-outs. Maximum stress at the fracture plane was also decreased from 3.7 MPa to 1.8 MPa, which indicates improved fixation after CSC augmentation. These results indicate that calcium sulfate cement is effective in augmenting fixation in osteoporotic bone. Calcium sulfate cement has potential application in the treatment of osteoporotic fractures.
What problem does this paper attempt to address?