Study on the Factors That Affect the Results of Critical Wavelength Determination

李竹,苏瑾,周世伟,宋伟民
DOI: https://doi.org/10.3969/j.issn.1006-3617.2008.06.009
2008-01-01
Abstract:[Objective]To investigate the characteristic of critical wavelength and study the appropriate test conditions by Labsphere UV-1000 Transmittance Analyzer.[Methods] The dependent factors such as substrate,temperature,humidity, balance time,blank control and the amount of sample adding were studied by Labsphere UV-1000 Transmittance Analyzer. [Results] Critical wavelength was independent on temperature and humidity and balance time. When the Labsphere UV-1000 Transmittance Analyzer is operated,different blank control did not affect the value of critical wavelength. But critical wavelength was incompatible due to different substrates and different applied amount.[Conclusion] Critical wavelength was stable and much less sensitive to unavoidable variations in operation. The appropriate test conditions for Labsphere UV-1000 Transmittance Analyzer are roughened quartz plates,application rate of 0.75 mg/cm2,balance time of 30 min,black roughened quartz plates as control,and usual room temperature and humidity.