Determination of the Contents of Water-soluble Vitamin for Injection by HPLC
SHI Hui,QIU Li-yan,JIN Yi,JIN Long-yu
DOI: https://doi.org/10.3321/j.issn:1001-2494.2007.19.019
2007-01-01
Abstract:OBJECTIVE To determine the contents of Water-soluble Vitamin for Injection by HPCE.METHODS The separation was performed on a fused-silica capillary column:67 cm(efficiency length 55 cm×50 μm),The mobile phase consisted of 50 mmol·L-1 SDS——50 mmol·L-1 Na2B4O7·10H2O solution(pH=8.33) eletroinjection:6.89 KPa×10 s;with running voltage of 15 kV.The detection wavelength was 214 nm.The column temperature was 25 ℃.RESULTS The linear ranges(n=5) for the determination of nicotinamide,Vitamin B6,D-biotin,Riboflavin 5′-phosphate sodium,Vitamin B12,Vitamin C,D-pantothenic acid sodium salt,folic acid and Vitamin B1 were 4.16~520.0 mg·L-1(r=0.999 1),0.40~50.0 mg·L-1(r=0.999 7),0.42~52.0 mg·L-1(r=0.999 3),0.40~50.0 mg·L-1(r=0.998 2),0.39~49.0 mg·L-1(r=0.999 6),8.00~1 000.0 mg·L-1(r=0.992 0),0.83~104.0 mg·L-1(r=0.999 5),0.42~53.0 mg·L-1(r=0.999 0),0.41~51.0 mg·L-1(r=0.999 8),respectively.The precision was good(with RSDs of 1.75%,1.04%,1.6%,2.84%,1.85%,3.26%,1.58%,2.5% and 1.92%,respectively(n=5).The detection limits were 0.60,0.16,0.24,0.23,0.21,0.45,0.41,0.20 and 0.18 mg·L-1,respectively.CONCLUSION The method of determining contents of water-soluble vitamin for injection by HPCE is easy to operate and the results of the determination are accurate.