Quantitative Detection of Cytomegalovirus DNA by Real-Time PCR in Transplant Recipients

ZHAO Xiao-tao,ZHANG Zheng,ZHANG Yan-rong,LIU Da-min,SUN Yuan-yuan
DOI: https://doi.org/10.3760/j:issn:1009-9158.2006.05.017
2006-01-01
Abstract:Objective To evaluate the advantages of cytomegalovirus (CMV) real-time PCR, to monitor CMV infection in a population of transplantation recipients in comparison with the qualitative PCR method.Methods 150 plasma samples and leukocytes samples collected from 59 bone marrow transplant recipients and 9 liver transplant recipients were tested by qualitative CMV PCR assay in parallel; Real-time CMV PCR using Roche Light Cycler was performed with 150 plasma samples and 54 control plasma samples.Results Using real-time PCR as the reference standard, the sensitivity and specificity of the qualitative CMV PCR assay with leukocytes samples were 62.5% and 95.5%, with plasma samples were 57.5% and 99.1% respectively. The recipients with the level of CMV DNA over 5×10~3 copies/ml by real-time PCR had higher percentage of developing CMV disease than those below 10~3 copies/ml.Conclusion The quantitative detection of CMV DNA by real-time PCR with plasma is a rapid, specific and sensitive method to monitor CMV infection in patients after transplantation and to guide antivirus therapy.
What problem does this paper attempt to address?