Ultratrace voltammetric method for the detection of DNA sequence related to human immunodeficiency virus type 1

Ruibing Wang,Chenghu Xue,Mei Gao,Honglan Qi,Chengxiao Zhang
DOI: https://doi.org/10.1007/s00604-010-0490-2
IF: 5.7
2010-01-01
Microchimica Acta
Abstract:An ultra-trace voltammetric method was developed for the determination of single strand DNA (ss-DNA) related to the human immunodeficiency virus type 1 (HIV-1). It is based on the signal amplification of carbon nanotubes loaded with silver nanoparticles and placed on a gold microelectrode. The capture ss-DNA (a 21-mer) possessing a thiol group at the 3′ end was self-assembled onto the surface of the gold microelectrode. It was then hybridized with target HIV-1 ss-DNA (a 42-mer) and further hybridized with the electrochemical probe (a 18-mer ss-DNA) tagged with multiwall carbon nanotubes and loaded with silver nanoparticles. The resulting formation of a DNA sandwich conjugate led to a strong electrochemical oxidation signal that was linearly proportional to the concentration of HIV-1 ss-DNA in the range from 1.0 to 100 pM. The detection limit was 0.5 pM (at an S / N of 3). This was equivalent to 0.05 fmol of HIV-1 ss-DNA in a volume of 20 μL. The relative standard deviation was 4.0% at 1.0 pM ( n = 11). Non-complementary ss-DNA of HIV-1 ss-DNA was effectively discriminated. This work demonstrates that the employment of the microelectrode and a sandwich hybridization model is promising in terms of sensitive and selective electrochemical detection of DNA. Figure Schematic diagram of the sandwich electrochemical detection for DNA hybridization
What problem does this paper attempt to address?