Effects of X-ray Scatter on Micro-Ct

GUO Xiaolian,TANG Zhiwei,ZHANG Hui,HU Guangshu
DOI: https://doi.org/10.16511/j.cnki.qhdxxb.2010.03.033
2010-01-01
Abstract:This study evaluates the magnitude and the effects of scatter of low kV X-ray in cone-beam micro-CT(micro-computed tomography) system for small animals.The magnitude of scatter is based on scatter-to-primary ratio(SPR) calculated using the beam stop method and estimated by polynormail-fitting extrapolation to 0 mm size stoppers.Lead foils with various slit widths were used to change the cone angle for the micro-CT imaging and its effect of the X-ray scattering.The standard phantom consisted of a uniform resin(40 mm diameter) representing a small mouse.The SPR at the center of the cone-beam system is 4.78% at the cone angle of 0.247° and is 27.58% at the cone angle of 5.67°.The effects of X-ray scattering on the reconstructed image uniformity and contrast ratio are also evaluated.