Development And Clinical Application Of Biosensor For Detecting Uric Acid In Whole-Blood

Chen Zhen-Cheng,Wang Hong-Yan,He Ji-Shan,Deng Zhen-Sheng
DOI: https://doi.org/10.3321/j.issn:0253-3820.2007.04.034
IF: 1.193
2007-01-01
Chinese Journal of Analytical Chemistry
Abstract:An amperometric uric acid biosensor was developed by immobilizing uric acid oxidase on carbon electrodes, in which, potassium ferricyanide was used to electron medium of chemistry response. The uric acid measured was oxidized and potassium ferricyanide was deacidized on the biosensor through catalysis of uric acid oxidase and supply of stationary 0.3 volt. As a result, the peak of oxidation-reduction was produced on the biosensor. The concentration of uric acid in product was measured indirectly by amperometry. The range of measuring uric acid was about 25-200 mg/L from the clinical experiment. It has better linear relationship with correlation coefficient of 0.9987. It reaches to steady value of currents in 10 seconds. It can be used for the analysis of blood with higher precision, the RSD is less than 2%.
What problem does this paper attempt to address?