Quantitative Determination of Acyclovir in Aqueous Humor by LC-MS

Liang Jin,Gang Wei,Wei-Yue Lu,Ling-Jie Xu,Jun Pan
DOI: https://doi.org/10.1365/s10337-006-0732-y
2006-01-01
Chromatographia
Abstract:A specific, sensitive and precise liquid chromatographic assay method was established using LC-MS for the determination of acyclovir (ACV) in aqueous humor (AH), which was directly injected onto an Inertsil ODS-3 C 18 column without any pretreatment. The Agilent 1100 series LC–MS system was operated under the electrospray ionization mode (ESI). The analyte was separated from endogenous substances with a mobile phase of methanol: water: acetic acid (5:95:0.1, v / v ) at a flow-rate of 0.3mL min −1 . A linear response was observed over the concentration range from 5 to 50ng mL −1 ( r =0.9993). Intra- and inter-day coefficients of variation were in the ranges 5.2–9.0% and 5.8–8.2%, respectively. The recovery was > 91.0%, and the limit of detection was approximate 1ng mL −1 . The pharmacokinetics of topically applied eye-drop and thermosetting gel were compared in rabbits utilizing the present method, the results demonstrated that LC-MS was a powerful tool for the detection of ACV in AH.