Research on accuracy of finger blood glucose level and synchronous venous plasma glucose level

TANG Jiaping,LU Huijuan,YU Haoyong,XIE Jun,YUAN Qiaoying,JIA Weiping
2010-01-01
Abstract:Objective To validate the accuracy of finger blood glucose test result by comparing the results of finger blood sample and venous blood sample,so as to provide evidence for self monitoring of blood glucose(SMBG) and blood glucose management in diabetes patients.Methods A total of 108 diabetic patients,who visited out endocrinology outpatient department,were randomly selected.The finger tip capillary blood and synchronous venous blood samples were collected under fasting condition and 2h postprandial using 3 glucose meters of different brands.Results The fasting plasma glucose was(6.28±1.76) mmol/L and the 2 h postprandial plasma glucose was(9.69±4.43) mmol/L in the venous blood samples.The fasting and 2 h postprandial finger blood glucose levels by 3 different meters were(6.53±1.56),(10.41±4.79) mmol/L;(6.28±1.44),(9.49±4.0) mmol/L;and(6.49±1.44),(9.88±3.96) mmol/L,respectively.The finger tip blood glucose values were only slightly higher than the venous plasma glucose values(all P>0.05).The coefficients of the 3 levels detected by 3 different meters with the venous plasma glucose values were 0.959,0.974,and 0.962,respectively.Conclusion Finger tip glucose detection can be used as a satisfactory method for diabetes patients for self-monitoring and management of their blood glucose.(Shanghai Med J,2010,33: 570-571)